4. Instrument Check
1 min
see the procedure in the service manual (sm 6008 5772 chapter 3 3 6) no action ok 1 check diagnostic center for any detected events (no events accepted) ☐ 2 check (view log) and create an "event log report" in instrument performance reports ☐ 3 perform "back up the entire database" ☐ 4 perform "clean up data" ☐ 5 redo "back up the entire database" ☐ 6 check "flow system check" (no warnings/errors accepted) ☐ 7 check "sample pipette leak check" (no warnings/errors accepted) ☐ 8 check "ir system check measure with no cuvette flow" (no warnings/errors accepted) ☐ 9 check "flow system leak diagnostic tool" (no warnings/errors accepted) ☐ 10 export wear log at "save service data to file" send to global help desk ☐ time required 0 5 hour table 4 instrument check